(!)Due to Microsoft's end of support for Internet Explorer 11 on 15/06/2022, this site does not support the recommended environment.
Part Number
Configured Part Number is shown.
E5CD-RX2A6M-002
![Temperature Controller (Digital Controller) [E5CD/E5CD-B] E5CD Outline Drawing](http://th.misumi-ec.com/en/linked/material/mech/OMR1/PHOTO/221302214995_004.jpg)
![Temperature Controller (Digital Controller) [E5CD/E5CD-B] E5CD-B Outline Drawing](http://th.misumi-ec.com/en/linked/material/mech/OMR1/PHOTO/221302214995_003.jpg)
![Temperature Controller (Digital Controller) [E5CD/E5CD-B] Parts Name Diagram](http://th.misumi-ec.com/en/linked/material/mech/OMR1/PHOTO/221302214995_002.jpg)
Temperature Controllers Rating
| Power supply voltage | Power supply voltage Type A: AC100-240V 50/60Hz Power supply voltage Type D: AC24V 50/60Hz /DC24V | |
|---|---|---|
| Allowable voltage fluctuation range | 85 to 110% of power supply voltage | |
| Power Consumption | Option -000 type: 5.2VA or less (AC100-240V), 3.1VA or less (AC24V) / 1.6W or less (DC24V) Other than the above: 6.5VA or less (AC100-240V), 4.1VA or less (AC24V) / 2.3W or less (DC24V) | |
| Sensor Input | Temperature Controllers input Thermocouple: K, J, T, E, L, U, N, R, S, B, C/W, PLII Platinum resistance thermometer: Pt100, JPt100 Non-contact Temperature Controllers sensor (ES1B): 10 to 70°C, 60 to 120°C, 115 to 165°C, 140 to 260°C Analog input Current input: 4 to 20mA, 0 to 20mA Voltage input: 1 to 5V, 0 to 5V, 0 to 10V | |
| Input impedance | Current input: 150 Ω or less, voltage input: 1 MΩ or more (When connecting to ES2-HB-N/THB-N, use a 1:1 connection.) | |
| Temperature Controllers Method | ON/OFF or 2-degree-of-freedom PID (with autotuning) | |
| Temperature Controllers Output | Relay Output | 1a AC250V 3A (resistive load) Electrical life 100,000 times Minimum applicable load 5V 10mA (reference value) |
| Voltage output (for driving SSR) | Output voltage DC12V±20% (PNP) Maximum load current 21mA, with short circuit protection circuit | |
| Linear Current Output | DC4~20mA/DC0~20mA Load: 500Ω or less Resolution: Approx. 10,000 | |
| Auxiliary Output | Score | 2 points |
| Output Specifications | Relay output 1a AC250V, E5CD 3A (resistive load), E5CD-B 2A (resistive load) Electrical life 100,000 times Minimum applicable load 5V 10mA (reference value) | |
| Event Input | Score | 2 points |
| External contact input specifications | During contact input: ON: 1kΩ or less OFF: 100kΩ or more | |
| Non-contact input: ON: Residual voltage 1.5V or less OFF: Leakage current 0.1mA or less | ||
| Current output: Approx. 7mA (per contact) | ||
| Transmission Output | Score | 1 point (depending on the model): Type with transmission output |
| Output Specifications | Current output: DC4-20mA Load: 500Ω Resolution: Approx. 10,000 Linear voltage output: DC1-5V Load: 1kΩ or more Resolution: Approx. 10,000 | |
| Setting method | Digital setting by front keys | |
| Instruction method | 11-segment digital display, single-light-emitting display, bar display character height PV: 14.9 mm, SV: 7.1 mm | |
| Multi-SP function* | Up to eight target values (SP0 to SP7) can be stored and selected by event input, key operation, or serial communication | |
| Bank switching function | none | |
| Other features | Adaptive Temperature Controllers, automatic filter adjustment function, water-cooled output adjustment function, predictive data (power-on time monitor, ambient temperature monitor, control output ON/OFF count monitor), parameter mask function, operation specification function at power-on, manual output, heating/cooling control, loop break alarm function, SP lamp, alarm function, heater break detection function (including SSR failure), 40% AT, 100% AT, manipulated variable limit, input digital filter, robust tuning, PV input correction, run/stop, protect, square root calculation function, manipulated variable change rate limit, simple calculation, temperature status display function, simple program function, input moving average, display brightness setting | |
| Ambient Temperature Controllers | -10 to +55°C (but no condensation or freezing) / 3-year warranty: -10 to +50°C (but no condensation or freezing) when installed as a standard unit | |
| Ambient humidity | Relative humidity 25-85% | |
| Storage temperature | -25 to +65°C (but no condensation or freezing) | |
| Altitude | Under 2,000m | |
| Recommended fuse | T2A, AC250V Time-lag low breaking capacity | |
| Installation environment | Overvoltage Category II, Pollution Degree 2 (EN/IEC/UL 61010-1) | |
* If you select from event input, the maximum number of target values is 4.
Temperature Controllers Performance
| Indication accuracy (ambient temperature 23℃) | Thermocouple: (±0.3% of indicated value or ±1°C, whichever is greater) ±1 digit or less *1 Platinum resistance thermometer: (±0.2% of indicated value or ±0.8°C, whichever is greater) ±1 digit or less Analog input: ±0.2%FS ±1 digit or less CT input: ±5%FS ±1 digit or less | |
|---|---|---|
| Transmission Output Accuracy | ±0.3%FS or less | |
| Temperature Controllers effect *2 | Thermocouple input (R, S, B, C/W, PLII): (±1% of indicated value or ±10°C, whichever is greater) ±1 digit or less Other thermocouple input: (±1% of indicated value or ±4°C, whichever is greater) ±1 digit or less *3 Platinum resistance thermometer input: (±1% of indicated value or ±2°C, whichever is greater) ±1 digit or less Analog input: ±1%FS ±1 digit or less CT input: ±5%FS ±1 digit or less | |
| Voltage effect *2 | ||
| Effects of electromagnetic interference (according to EN 61326-1) | ||
| Input Sampling Period | 50ms | |
| Adjustment Sensitivity | Temperature input: 0.1 to 999.9°C/°F (in 0.1°C/°F increments) Analog input: 0.01 to 99.99%FS (in 0.01%FS increments) | |
| Proportional band (P) | Temperature Controllers input: 0.1 to 999.9°C/°F (in 0.1°C/°F units) Analog input: 0.1 to 999.9%FS (in 0.1%FS units) | |
| Integration time (I) | 0 to 9999s (in 1s units), 0.0 to 999.9s (in 0.1s units) *4 | |
| Differential time (D) | 0 to 9999s (in 1s units), 0.0 to 999.9s (in 0.1s units) *4 | |
| Cooling proportional band (P) | Temperature Controllers input: 0.1 to 999.9°C/°F (in 0.1°C/°F units) Analog input: 0.1 to 999.9%FS (in 0.1%FS units) | |
| Cooling integration time (I) | 0 to 9999s (in 1s units), 0.0 to 999.9s (in 0.1s units) *4 | |
| Cooling differential time (D) | 0 to 9999s (in 1s units), 0.0 to 999.9s (in 0.1s units) *4 | |
Dedicated to adaptive control | Proportional band for target response (P) | Temperature Controllers input: 0.1 to 999.9°C/°F (in 0.1°C/°F increments) |
| Integral time for target response (I) | 0 to 9999s (in 1s units), 0.0 to 999.9s (in 0.1s units) *4 | |
| Derivative time for target response (D) | 0 to 9999s (in 1s units), 0.0 to 999.9s (in 0.1s units) *4 | |
Proportional band for disturbance (P) | Temperature Controllers input: 0.1 to 999.9°C/°F (in 0.1°C/°F increments) | |
| Disturbance integral time (I) | 0 to 9999s (in 1s units), 0.0 to 999.9s (in 0.1s units) *4 | |
| Disturbance Differential Time (D) | 0 to 9999s (in 1s units), 0.0 to 999.9s (in 0.1s units) *4 | |
| Control Period | 0.1, 0.2, 0.5, 1 to 99 s (in 1 s increments) | |
| Manual Reset Value | 0.0 to 100.0% (in 0.1% increments) | |
| Alarm Setting Range | -1999 to 9999 (decimal point position depends on input type) | |
| Effect of signal source resistance | Thermocouple: 0.1℃/Ω or less (100Ω or less), Platinum resistance thermometer: 0.1℃/Ω or less (10Ω or less) | |
| Insulation resistance | 20MΩ or more (500V DC applied) | |
| Withstand voltage | AC3,000V 50 or 60Hz 1min (opposite polarity live terminals) | |
| vibration | malfunction | 10~55Hz 20m/s2 3 axis direction 10min |
| Durability | 10~55Hz 20m/s2 3 axial directions 2h | |
| impact | malfunction | 100m/s23 axial directions, 3 times each |
| Durability | 300m/s23 axial directions, 3 times each | |
| mass | Main unit: Approx. 120g Mounting adapter: Approx. 10g | |
| Protective structure | Front: IP66, Rear case: IP20, Terminal: IP00 | |
| Memory Protection | Non-volatile memory (number of writes: 1 million times) | |
| Configuration Tool | E5CD: CX-Thermo Ver.4.66 or higher E5CD-B: CX-Thermo Ver.4.67 or higher | |
| Configuration tool port | Top of E5CD/E5CD-B: Use the E58-CIFQ2 USB-serial conversion cable to connect to the USB port on the computer. *5 | |
| standard | Certification Standards | cULus: UL 61010-1/CSA C22.2 No.61010-1, Korea Wireless Regulations (Radio Act: KC Mark) |
| Compliance standards | EN 61010-1 (IEC 61010-1), RCM standard | |
| EMC Directive | EMI EN 61326-1 *6 Radiated interference field strength EN 55011 Group1 class A Noise terminal voltage EN 55011 Group1 class A EMS EN 61326-1 *6 Electrostatic discharge immunity EN 61000-4-2 Electromagnetic field immunity EN 61000-4-3 Burst noise immunity EN 61000-4-4 Conducted interference immunity EN 61000-4-6 Surge immunity EN 61000-4-5 Voltage dip/interruption immunity EN 61000-4-11 | |
*1. K (-200 to 1300°C range), T, N below -100°C, and U, L are within ±2°C ±1 digit. B below 400°C is not specified. B from 400 to 800°C is within ±3°C.
R, S below 200°C is within ±3°C ±1 digit. C/W is (±0.3% PV or ±3°C, whichever is greater) ±1 digit.
PLII is (±0.3% PV or ±2°C, whichever is greater) ±1 digit.
*2. Conditions: Ambient Temperature Controllers: -10°C to 23°C to 55°C Voltage range: -15 to +10% of rated voltage
*3. K sensor below -100°C is within ±10°C
*4. The unit depends on the "Integral/Differential time unit" setting.
*5. External serial communication (RS-485) and USB-serial conversion cable communication can be used simultaneously.
*6. Industrial electromagnetic environment (EN/IEC 61326-1 Table 2)
| Part Number |
|---|
| E5CD-RX2A6M-002 |
| Part Number | Standard Unit Price | Minimum order quantity | Volume Discount | Days to Ship | Output Type | Failure Detecting Function | Power Supply Voltage (V) | Control Output 1 | Terminal shape | Event Input (Point) | Heater Breakage, SSR Fault Detection Capability | Communication | Transmission Output |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
MYR 1,480.11 | 1 Piece(s) | 5 Day(s) or more | Relay contact | Heater Breakage / SSR Breakdown | 100 to 240 AC | Relay output | Screw terminal block type | - | 1 point | RS-485 | - |
Loading...
| Input Method | Thermocouple / Platinum temperature measuring resistor / Non-Contact Temperature Sensor / Voltage / Current | Control Method | ON/OFF Control / PID control with auto tuning | Number of Control Points | 1 |
|---|---|---|---|---|---|
| Setting Method | Key input | Control Output 2 | NA | Auxiliary Output Points | 2 independent |
| Specifications | 48×48 mm | Input type | Full multi input |
Please check the type/dimensions/specifications of the part E5CD-RX2A6M-002 in the Temperature Controller (Digital Controller) [E5CD/E5CD-B] series.
| Part Number |
|---|
| E5CD-CX2ABM-000 |
| E5CD-CX2ABM-004 |
| E5CD-CX2D6M-000 |
| E5CD-RX2ABM-000 |
| E5CD-RX2ABM-002 |
| E5CD-RX2D6M-000 |
| Part Number | Standard Unit Price | Minimum order quantity | Volume Discount | Days to Ship | Output Type | Failure Detecting Function | Power Supply Voltage (V) | Control Output 1 | Terminal shape | Event Input (Point) | Heater Breakage, SSR Fault Detection Capability | Communication | Transmission Output |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
MYR 1,153.29 | 1 Piece(s) | 9 Day(s) or more | Current | No | 100 to 240 AC | - | Push-in plus terminal block type | - | - | - | - | ||
MYR 1,637.22 | 1 Piece(s) | 9 Day(s) or more | Current | No | 100 to 240 AC | - | Push-in plus terminal block type | 2 | - | RS-485 | - | ||
MYR 1,075.24 | 1 Piece(s) | 9 Day(s) or more | Current | No | 24 AC/DC | - | Screw terminal block type | - | - | - | - | ||
MYR 1,008.78 | 1 Piece(s) | 9 Day(s) or more | Relay contact | No | 100 to 240 AC | - | Push-in plus terminal block type | - | - | - | - | ||
MYR 1,542.98 | 1 Piece(s) | 9 Day(s) or more | Relay contact | SSR Breakdown / Heater Breakage | 100 to 240 AC | - | Push-in plus terminal block type | - | 1 point | RS-485 | - | ||
MYR 958.57 | 1 Piece(s) | 9 Day(s) or more | Relay contact | No | 24 AC/DC | Relay output | Screw terminal block type | - | - | - | - |
How can we improve?
How can we improve?
While we are not able to respond directly to comments submitted in this form, the information will be reviewed for future improvement.
Customer Privacy Policy
Thank you for your cooperation.
While we are not able to respond directly to comments submitted in this form, the information will be reviewed for future improvement.
Please use the inquiry form.
Customer Privacy Policy