Part Number
Configured Part Number is shown.
ES1-LP10-N


(The ES1-LW50-N does not have the laser output port (6) or laser switch (7).)

(The ES1-LW100-N does not have the laser output port (6) or laser switch (7).)


ES1-N
| type | No aiming laser | With aiming laser | |||||
|---|---|---|---|---|---|---|---|
| Model number | Model ES1-LP3-N | Model ES1-LP10-N | Model ES1-LW50-N | Model ES1-LW100-N | Model ES1-LW50L-N | Model ES1-LW100L-N | |
| Detector element | Thermopile | ||||||
| lens | silicon | ||||||
| Wavelength | 8–14 μm | ||||||
| output | USB Output | Resolution: 0.1°C | |||||
| Current Output | 4~20mA/-50~500℃*1 Load impedance 250Ω or less Resolution 0.24μA | 4~20mA/-50~1,000℃*1 Load impedance 250Ω or less Resolution 0.24μA | 4~20mA/-50~500℃*1 Load impedance 250Ω or less Resolution 0.24μA | 4~20mA/-50~1,000℃ *1 Load impedance 250Ω or less Resolution 0.24μA | |||
| Power supply voltage | DC12 to 24V/USB bus power (USB bus power only, other than current output possible) | ||||||
| Allowable power supply voltage fluctuation | 95 to 105% of power supply voltage | ||||||
| Current consumption | 30mA or less (24V DC) | ・30mA or less (DC24V, when aiming laser is not used) ・40mA or less (DC24V, when aiming laser is used) | |||||
| Cable length | 2m±5cm | ||||||
| Measurement accuracy | USB output : Within ±(-8%rdg+1)℃ (measurement temperature: -50 to 0℃) Within ±1℃ (measurement temperature: 0 to 200℃) Within ±0.5%rdg℃ (200 to 500℃) Current output : Difference from USB output is within (±0.1% output range)℃ | USB output : Within ±(-8%rdg+1)℃ (measurement temperature: -50 to 0℃) Within ±1℃ (measurement temperature: 0 to 200℃) Within ±0.5%rdg℃ (200 to 1,000℃) Current output : Difference from USB output is within (±0.1% output range)℃ | USB output : Within ±(-8%rdg+1)℃ (measurement temperature: -50 to 0℃) Within ±1℃ (measurement temperature: 0 to 200℃) Within ±0.5%rdg℃ (200 to 500℃) Current output : Difference from USB output is within (±0.1% output range)℃ | USB output : Within ±(-8%rdg+1)℃ (measurement temperature: -50 to 0℃) Within ±1℃ (measurement temperature: 0 to 200℃) Within ±0.5%rdg℃ (200 to 1,000℃) Current output : Difference from USB output is within (±0.1% output range)℃ | |||
| Reproducibility | ・Within 0.5℃ | ・Within 1℃ (measurement temperature: -50 to 0℃) ・Within 0.5℃ (measurement temperature: 0 to 500℃) ・Within 1℃ (measurement temperature: 500 to 1,000℃) | ・Within 0.5℃ | ・Within 1℃ (measurement temperature: -50 to 0℃) ・Within 0.5℃ (measurement temperature: 0 to 500℃) ・Within 1℃ (measurement temperature: 500 to 1,000℃) | |||
| Temperature Drift | ・Within ±0.5℃/℃ (measurement temperature: -50 to 0℃) ・Within ±0.25℃/℃ (measurement temperature: 0 to 500℃) | ・Within ±0.5℃/℃ (measurement temperature: -50 to 0℃) ・Within ±0.25℃/℃ (measurement temperature: 0 to 500℃) ・Within ±0.5℃/℃ (measurement temperature: 500 to 1,000℃) | ・Within ±0.5℃/℃ (measurement temperature: -50 to 0℃) ・Within ±0.25℃/℃ (measurement temperature: 0 to 500℃) | ・Within ±0.5℃/℃ (measurement temperature: -50 to 0℃) ・Within ±0.25℃/℃ (measurement temperature: 0 to 500℃) ・Within ±0.5℃/℃ (measurement temperature: 500 to 1,000℃) | |||
| Effects of electromagnetic interference *2 | Electromagnetic field discharge immunity | Within ±10℃ (80MHz to 1.0GHz) | |||||
| Conducted interference immunity | Within ±10℃ (150kHz to 80MHz) | ||||||
| Response Time | Current output: 0.14 s or less (95% response, moving average 1) | ||||||
| Emissivity | Factory setting: 0.95 Can be changed from 0.100 to 1.999 using the setting tool | ||||||
| Moving average function | Factory setting: 10 Can be changed from 1 to 1,000 using the setting tool | Factory setting: 50 Can be changed from 1 to 1,000 using the setting tool | Factory setting: 10 Can be changed from 1 to 1,000 using the setting tool | Factory setting: 50 Can be changed from 1 to 1,000 using the setting tool | |||
| Operating temperature and humidity range | Temperature: 0 to 55°C, Humidity: 35 to 85% (no condensation) | ||||||
| Storage Temperature Range | -20 to 55°C (no condensation) | ||||||
| Vibration resistance | malfunction | 10~55Hz 20m/s2 3 axis direction 10min | |||||
| Durability | 10 to 55Hz, half amplitude 0.75mm, 3 axial directions 2h | ||||||
| Protective structure | No | ||||||
| Compliance standards | CE, KC, FCC, RCM | ||||||
| - | IEC60825-1, PSC, FDA | ||||||
| External dimensions | W32mm×H40mm×L58mm | W32mm×H40mm×L90.3mm | W32mm×H40mm×L60.4mm | W32mm×H40mm×L92.7mm | |||
| mass | Approx. 95g | Approx. 115g | Approx. 95g | Approx. 115g | |||
| Standard accessories | Instruction manual, mounting bracket, power/output cable (2m), installation gauge | Instruction manual, mounting bracket, power/output cable (2m) | |||||
*1. The default range is 0 to +500°C. When using in -50°C to +500°C or -50°C to +1,000°C, the range must be changed using the ES1-TOOLS.
*2. Industrial electromagnetic environment (EN/IEC 61326-1 Table 2). In this case, the measurement error is ±10°C.
Model ES1B
| item | Model number | |
|---|---|---|
| Model ES1B | ||
| Power supply voltage | DC12/24V | |
| Allowable power supply voltage fluctuation | 90% to 110% of power supply voltage | |
| Current consumption | 20mA or less | |
| Measurement Temperature Range | 10~70℃, 60~120℃, 115~165℃, 140~260℃ | |
| Accuracy*1 | ±5℃*2 | ±2%PV or ±2℃, whichever is greater |
| ±10℃*2 | ±4%PV or ±4℃, whichever is greater | |
| ±30℃*2 | ±6%PV or ±6℃, whichever is greater | |
| ±40℃*2 | ±8%PV or ±8℃, whichever is greater | |
| Reproducibility | ±1%PV or ±1℃, whichever is greater | |
| Temperature Drift | ±0.4℃/℃ or less | |
| Effects of electromagnetic interference | Field Strength Immunity | Within ±10℃ (80-800MHz) |
| Within ±30℃ (800-1000MHz) | ||
| Conducted interference immunity | ±10℃ | |
| Measurement distance: Field of view diameter | 1:1 typ. | |
| Wavelength | 6.5 to 14.0 μm | |
| Light receiving element | Thermopile | |
| Response speed | Approximately 300 ms at 63% response | |
| Output Impedance | 1 to 4 kΩ | |
| Operating Temperature Range | -25 to +70°C (no condensation or freezing) | |
| Permitted ambient humidity | Relative humidity 35-85% | |
| Vibration resistance (durability) | 98m/s2 10~55Hz 2h each in X, Y, and Z directions | |
| Impact resistance (durability) | 300m/s2 3 times each in X, Y, and Z directions | |
| Body Material | ABS resin | |
| Protective structure | IP65 | |
| Compliance standards | CE marking*3 | |
| mass | Approx. 120g | |
| Lead Wire | Compensating wire 3m | |
| PVC coating, heat resistant up to 70℃ | ||
*1.For K thermocouple characteristics and emissivity of 0.98
*2.Example of temperature change from any reference temperature of the measurement object
: When corrected to a reference temperature of 50°C, the accuracy is
the greater of ±2% PV or ±2°C at 55°C, and ±4% PV or ±4°C at 60°C.*3.EN61326-1: Industrial electromagnetic environment (EN/IEC61326-1 Table 2)
ES1C
| item | Model number | |
|---|---|---|
| Model ES1C | ||
| Power supply voltage | DC12-24V | |
| Allowable voltage fluctuation range | 90 to 110% of rated voltage | |
| Current consumption | 70mA or less | |
| Measurement Temperature Range | 0~400℃ | |
| Measurement accuracy | 0 to 200℃: ±2℃ 201 to 400℃: ±1% PV (emissivity 0.95) | |
| Effects of electromagnetic interference | Field Strength Immunity | ±10℃ max |
| Conducted interference immunity | ±10℃ max | |
| Response Time | 100ms/90% | |
| Reproducibility | ±1% of reading | |
| Wavelength | 8–14 μm | |
| Light receiving element | Thermopile | |
| Emissivity | Fixed at 0.95 | |
| Current Output | DC4 to 20mA Load 250Ω or less | |
| Ambient Temperature Range | During use: 0 to +70°C During storage: -20 to +70°C (no freezing or condensation) | |
| Ambient Humidity Range | Use/storage: Relative humidity 35-85% | |
| Vibration resistance (durability) | 10~55Hz, amplitude 1.5mm, 2h each in X, Y, and Z directions | |
| mass | 180g | |
| Protective structure | IP67 equivalent | |
| Compliance standards | CE marking* | |
*EN61326-1: Industrial electromagnetic environment (EN/IEC61326-1 Table 2)
| No. | name |
|---|---|
| (1) | lens |
| (2) | Mounting bracket screw holes |
| (3) | Tripod screw hole |
| (4) | Power output connector |
| (5) | Micro USB Connector |
| (6) | Laser output port |
| (7) | Laser Switch |
| (8) | Installation gauge fixing hole |
| Part Number |
|---|
| ES1-LP10-N |
| Part Number | Standard Unit Price | Minimum order quantity | Volume Discount | Days to Ship | Type | Operating Environment | Operating Temperature Range (°C) | Lead Wire Length F (m) | Lead Wire Coating Material | Measurement temperature range (°C) | Specification | Mounting part shape |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
MYR 4,098.71 | 1 Piece(s) | 9 Day(s) or more | - | Standard | 0~55 | 2 | - | -50 to 500 | - | - |
Loading...
| Measurement Object | Temperature | Operating Humidity Range(%) | 35::85 | Power Supply Voltage(V) | DC12 to 24 |
|---|
How can we improve?
How can we improve?
While we are not able to respond directly to comments submitted in this form, the information will be reviewed for future improvement.
Customer Privacy Policy
Thank you for your cooperation.
While we are not able to respond directly to comments submitted in this form, the information will be reviewed for future improvement.
Please use the inquiry form.
Customer Privacy Policy